The accuracy of our Micro Ohmmeters is defined as ±(0,2% rdg + 0,2% FS) provided
the following two conditions are met:
- the measurement current
is at least 100A
- the measurement range
is between 100µΩ and 20mΩ
If the current is below 100A the accuracy can be established following
the table below:
| 50 A |
0,25%+1µΩ |
| 20 A |
0,5%+2µΩ |
| 10 A |
1%+5µΩ |
Measurement of specimens below 100µΩ is not recommended with current
below 100A.
For specimen with resistance between 20µΩ and 100µΩ the accuracy is
defined following the table below:
| 600 A |
0,2%+0,5µΩ |
| 500 A |
0,2%+0,6µΩ |
| 400 A |
0,2%+0,7µΩ |
| 300 A |
0,2%+0,8µΩ |
| 200 A |
0,2%+0,9µΩ |
| 100 A |
0,2%+1,0µΩ |
For specimen with resistance above 20mOhm accuracy is defined using table:
| 50 A |
0,25% |
| 20 A |
0,5% |
| 10 A |
1% |
If you need a higher accuracy, then you need to:
- verify instrument
calibration using provided Test Shunt before testing
- always use the max
current available for particular range*
- do two measurement
in two directions, the second measurement with inverted connection on
the VS input. Calculate the average of the two measurements.
*Max current for a given range depends on the supply voltage, length
and cross section of cables (cable resistance), and max measurement range
on the Voltage sense input. Max measurement range on the Voltage sense
input is 2,5V
In case the instrument cannot reach/generate pre-set test current, display
will show the value of the maximum current achieved and the result.
This result is of the lower accuracy, so we recommend that this test
is repeated pre-selecting the first lower current range.
Back to products
Back to support |